照射損傷評価のための遠隔操作式振動試料型磁力計の開発


英字タイトル:
Development of Remote Operated Vibrating Sample Magnetometerfor Evaluation of Irradiation Damage
著者:
高屋 茂 Shigeru TAKAYA 山県 一郎 Ichiro YAMAGATA 永江 勇二 Yuji NAGAE 若井 栄一 Eiichi WAKAI 青砥 紀身 Kazumi AOTO
発刊日:
公開日:
カテゴリ: 第6回
キーワードタグ:

概要

The remote operated vibrating sample magnetometer (VSM) for neutron irradiated samples has been developed. The maximum range of applied magnetic field is more t han £0.5/10 A/m, and the resolution of magnetization is higher than 5 x10-A'm2. The magnetization curves of neutron irradiated samples were measured by this VSM , and the relationships between dose, which is one of representative irradiation parameter, and magnetic properties such as magnetic coercive force. As result, it was revealed that there are good correlations between them. This fact leads the possibility of nondestructive evaluation of irradiation damage by using VSM,